13.5 Validation / GLP

796 Titroprocessor, Instructions for use

286

286

sion of the endpoint volume (

RE) against sample size

(

SS).

Sample size

a

sys

Standard limiting value

± 0.050 mL

Validation interval

Time interval for the validity of the validation. If a valida-
tion has expired a message will be shown each time the
Titroprocessor is switched on.
Standard limiting value

‘12 month(s)'

Press the [Sample silo] softkey on the

GLP main page to enter the

sample weight. The sample weights should be scattered at random within a
wide range. The sample silo can be stored and called up at a later date.

By pressing the [Start valid.] softkey the validation can be started.
You will be required to answer the following questions:

Do you really want to overwrite current method [y]

Do you really want to reset all statistics data [y]

Do you want to set statistics 'on' [y]

Do you want to set silo 'on' [y]

Do you want to set auto start 'on' [y]

Did you set up all necessary silo data [y]

Do you change samples manually [n]

(*)

Accept all previous settings [y]

When all standard settings have been confirmed by pressing the

key the validation can be carried out after you have entered your name as
operator.

Validation of xx-xx-xx done by user: [ ]