App index, Te sets sampling conditions for event data, Sz sets a zone – Yokogawa Value Series FX1000 User Manual
Page 99: Sp sets a partial expanded display
3-17
IM 04L21B01-17EN
Commands
1
2
3
4
5
6
App
Index
TE
Sets sampling conditions for
event data
Syntax
TE p1,p2,p3,p4,p5,p6
p1 1
p2 Sample rate (125MS, 250MS, 500MS,
1S, 2S, 5S, 10S, 30S, 1MIN, 2MIN, 5MIN,
10MIN)
p3 Sample mode
FREE
Starts data acquisition at memory
start and stops data acquisition at
memory stop.
SINGLETRIGGER
Acquires data once
for a specified length of time after
a trigger occurs and then stops.
REPEATTRIGGER
Acquires data and
then enters the trigger-wait state.
p4 Sampling time (10MIN, 20MIN, 30MIN, 1H,
2H, 3H, 4H, 6H, 8H, 12H, 1DAY, 2DAY,
3DAY, 5DAY, 7DAY, 10DAY, 14DAY, 31DAY)
p5 Pretrigger length as percentage (0, 5, 25,
50, 75, 95, 100)
p6 Key trigger source disable or enable (OFF,
ON)
Parameters p5 to p6 are valid when p3 is set
to SINGLETRIGGER or REPEATTRIGGER.
Query
TE[p1]?
Example Acquire data at a sampling rate of 125 ms for 10
minutes using single trigger mode.
TE1,125MS,SINGLETRIGGER,10MIN
Description • You cannot choose a sample interval that is
shorter than the scan interval.
• You cannot use this command to configure
settings while recording (memory sampling) is
in progress.
SZ
Sets a zone
Syntax
SZ p1,p2,p3
p1 Measurement or computation channel
number
p2 Lower zone boundary position (0 to 95) as a
percentage
p3 Upper zone boundary position (5 to 100) as
a percentage
Query
SZ[p1]?
Example Display channel 002 in a zone between 30% and
50%.
SZ002,30,50
Description • You can specify computation channels
(including in queries) on models with the /M1,
/PM1, or /PWR1 math option.
• Set the boundary positions as percentages
of the entire amplitude axis in the waveform
display area.
• The zone size must be at least 5%.
• Set the parameters so that the upper zone
boundary position is greater than the lower
zone boundary position.
SP
Sets a partial expanded display
Syntax
SP p1,p2,p3,p4
p1 Measurement or computation channel
number
p2 Partial expanded display (ON, OFF)
p3 Boundary position (1 to 99) as a percentage
p4 Boundary value
Query
SP[p1]?
Example Partially expand the display of channel 001. Set
the boundary position to 25% and the boundary
value to 1.00 V.
SP001,ON,25,100
Description • You can specify computation channels
(including in queries) on models with the /M1,
/PM1, or /PWR1 math option.
• For a channel whose input range is set to
SKIP (using the SR command), p2 cannot be
set to ON.
• For a channel whose computation channel is
turned off (using the SO command), p2 cannot
be set to ON.
• Set p3 as a percentage of the range defined
by the span upper and lower limits (scaling
upper and lower limits when scaling is
enabled).
• Set p4 to a value from (span upper limit – 1) to
(span lower limit + 1). If scaling is enabled, set
p4 to a value from (scaling upper limit – 1) to
(scaling lower limit + 1).
• The decimal place and the number of digits
are the same as those for the span or scaling
settings (see the SR command).
• You can use this command (including its
query) when the partial expanded display
function is set to ON (using the WU
command).
• You cannot use this command if the partial
expanded display range does not exist (for
example when the span range is 1).
• You cannot use the partial expanded display
on a channel that is set to Log scale (/LG1
option).
3.4 Setting Commands