beautypg.com

App index, Te sets sampling conditions for event data, Sz sets a zone – Yokogawa Value Series FX1000 User Manual

Page 99: Sp sets a partial expanded display

background image

3-17

IM 04L21B01-17EN

Commands

1

2

3

4

5

6

App

Index

TE

Sets sampling conditions for

event data

Syntax

TE p1,p2,p3,p4,p5,p6
p1 1
p2 Sample rate (125MS, 250MS, 500MS,

1S, 2S, 5S, 10S, 30S, 1MIN, 2MIN, 5MIN,
10MIN)

p3 Sample mode

FREE

Starts data acquisition at memory
start and stops data acquisition at
memory stop.

SINGLETRIGGER

Acquires data once

for a specified length of time after
a trigger occurs and then stops.

REPEATTRIGGER

Acquires data and

then enters the trigger-wait state.

p4 Sampling time (10MIN, 20MIN, 30MIN, 1H,

2H, 3H, 4H, 6H, 8H, 12H, 1DAY, 2DAY,
3DAY, 5DAY, 7DAY, 10DAY, 14DAY, 31DAY)

p5 Pretrigger length as percentage (0, 5, 25,

50, 75, 95, 100)

p6 Key trigger source disable or enable (OFF,

ON)

Parameters p5 to p6 are valid when p3 is set
to SINGLETRIGGER or REPEATTRIGGER.

Query

TE[p1]?

Example Acquire data at a sampling rate of 125 ms for 10

minutes using single trigger mode.
TE1,125MS,SINGLETRIGGER,10MIN

Description • You cannot choose a sample interval that is

shorter than the scan interval.

• You cannot use this command to configure

settings while recording (memory sampling) is
in progress.

SZ

Sets a zone

Syntax

SZ p1,p2,p3
p1 Measurement or computation channel

number

p2 Lower zone boundary position (0 to 95) as a

percentage

p3 Upper zone boundary position (5 to 100) as

a percentage

Query

SZ[p1]?

Example Display channel 002 in a zone between 30% and

50%.
SZ002,30,50

Description • You can specify computation channels

(including in queries) on models with the /M1,
/PM1, or /PWR1 math option.

• Set the boundary positions as percentages

of the entire amplitude axis in the waveform
display area.

• The zone size must be at least 5%.
• Set the parameters so that the upper zone

boundary position is greater than the lower
zone boundary position.

SP

Sets a partial expanded display

Syntax

SP p1,p2,p3,p4
p1 Measurement or computation channel

number

p2 Partial expanded display (ON, OFF)
p3 Boundary position (1 to 99) as a percentage
p4 Boundary value

Query

SP[p1]?

Example Partially expand the display of channel 001. Set

the boundary position to 25% and the boundary
value to 1.00 V.
SP001,ON,25,100

Description • You can specify computation channels

(including in queries) on models with the /M1,
/PM1, or /PWR1 math option.

• For a channel whose input range is set to

SKIP (using the SR command), p2 cannot be
set to ON.

• For a channel whose computation channel is

turned off (using the SO command), p2 cannot
be set to ON.

• Set p3 as a percentage of the range defined

by the span upper and lower limits (scaling
upper and lower limits when scaling is
enabled).

• Set p4 to a value from (span upper limit – 1) to

(span lower limit + 1). If scaling is enabled, set
p4 to a value from (scaling upper limit – 1) to
(scaling lower limit + 1).

• The decimal place and the number of digits

are the same as those for the span or scaling
settings (see the SR command).

• You can use this command (including its

query) when the partial expanded display
function is set to ON (using the WU
command).

• You cannot use this command if the partial

expanded display range does not exist (for
example when the span range is 1).

• You cannot use the partial expanded display

on a channel that is set to Log scale (/LG1
option).

3.4 Setting Commands