3-12
3.4 Setting Commands
IM 04L21B01-17EN
3.4 Setting Commands
SR
Sets an input range
You cannot use this command to configure
settings while recording (memory sampling) or
computation is in progress.
When Setting Channels to Skip
Syntax
SR p1,p2
p1 Measurement channel number
p2 Setting type (SKIP)
Query
SR[p1]?
Example Skip channel 001.
SR001,SKIP
Description • Channels set to SKIP are not measured.
• Set p1 by referring to the table in section 3.3.
When Setting the Channels to Voltage, TC, RTD, or
ON/OFF Input
Syntax
SR p1,p2,p3,p4,p5
p1 Measurement channel number
p2 Input type
VOLT
DC voltage
TC
Thermocouple
RTD
Resistance temperature detector
DI
ON/OFF input
p3 Measuring range
p4 Span lower limit
p5 Span upper limit
Query
SR[p1]?
Example Set the channel 001 input type to TC type R, the
span lower limit to 0°C, and the span upper limit
to 1760.0°C.
SR001,TC,R,0,17600
Description • Set p1 and p3 by referring to the table in
section 3.3.
• For parameters p4 and p5, enter values with
five digits or less excluding the decimal point.
When Computing the Difference between Channels
Syntax
SR p1,p2,p3,p4,p5,p6,p7
p1 Measurement channel number
p2 Setting type (DELTA)
p3 Input type
VOLT
DC voltage
TC
Thermocouple
RTD
Resistance temperature detector
DI
ON/OFF input
p4 Measuring range
p5 Span lower limit
p6 Span upper limit
p7 Reference channel number (measurement
channel number)
Query
SR[p1]?
Example Set the channel 010 setting type to differential
computation between channels with the
reference channel set to 001, and set the input
type to TC. Set the measuring range to R. Set the
span lower limit to 10.0°C and the span upper limit
to 100.0°C.
SR010,DELTA,TC,R,100,1000,001
Description • Set p1 and p4 by referring to the table in
section 3.3.
• For parameters p5 and p6, enter values with
five digits or less excluding the decimal point.
When Setting Channels to Scaling
Syntax
SR p1,p2,p3,p4,p5,p6,p7,p8,p9,p10
p1 Measurement channel number
p2 Setting type (SCALE)
p3 Input type
VOLT
DC voltage
TC
Thermocouple
RTD
Resistance temperature detector
DI
ON/OFF input
p4 Measuring range
p5 Span lower limit
p6 Span upper limit
p7 Scaling lower limit (–30000 to 30000)
p8 Scaling upper limit (–30000 to 30000)
p9 Scaling decimal place (0 to 4)
p10 Unit (up to 6 characters)
Query
SR[p1]?
Example Convert the DC voltage measured on channel
002 to DC current. Set the measuring range to
6 V, the span lower limit to 1 V, the span upper
limit to 5 V, the scaling lower limit to 1.00 A, and
the scaling upper limit to 5.00 A.
SR002,SCALE,VOLT,6V,1000,5000,100,
500,2,A
Description • Set p1 and p4 by referring to the table in
section 3.3.
• For parameters p5 and p6, enter values with
five digits or less excluding the decimal point.
• For parameters p7, p8, and p9, either set all
three parameters, or omit all three parameters.