Rn sets basic login, Rp sets user limitations – Yokogawa Value Series FX1000 User Manual
Page 120
3-38
IM 04L21B01-17EN
RN
Sets basic login
Syntax
RN p1,p2
p1 Auto logout (OFF, 1MIN, 2MIN, 5MIN,
10MIN)
p2 Operation without Login
OFF
Disables FX operation
DISPLAY Only enables screen operation
Query
RN?
Example Set the auto logout time to 1 minute, and disable
FX operation when logged out.
RN1MIN,OFF
RP
Sets user limitations
Syntax
RP p1,p2,...
p1 User limitation number (1 to 10)
p2 User limitation item (KEY, FUNC, MEDIA)
Parameter p3 and subsequent parameters vary
as shown below depending on the p2 setting.
When p2 is set to KEY
Syntax
p3 START key (FREE, LOCK)
p4 STOP key (FREE, LOCK)
p5 MENU key (FREE, LOCK)
p6 USER key (FREE, LOCK)
p7 DISP/ENTER key (FREE, LOCK)
When p2 is set to FUNC (function key)
Syntax
p3 Alarm ACK (FREE, LOCK)
p4 Message and batch (FREE, LOCK)
p5 Computation (FREE, LOCK)
p6 Data save (FREE, LOCK)
p7 E-mail and FTP (FREE, LOCK)
p8 Time operations (FREE, LOCK)
p9 Display operations (FREE, LOCK)
Description Even on models that do not have the Ethernet
interface (/C7 option), if you set p2 to FUNC, you
can specify a value for parameter p7 (the E-mail
and FTP setting), but it will be ignored.
When p2 is set to MEDIA (external storage media)
Syntax
p3 External storage media operations (FREE,
LOCK)
p4 Setting load operations (FREE, LOCK)
Description You cannot use this command with p2 set to
MEDIA on models that do not have a CF card
slot or USB interface (/USB1 option).
Query
RP[p1,[p2]]?
Example Lock the START, STOP, and DISP/ENTER keys.
RP1,KEY,LOCK,LOCK,,,LOCK
RO
Sets the type of report and when
to create reports (/M1, /PM1, and
/PWR1 options)
When you are not creating reports
Syntax
RO p1
p1 Report type (OFF)
Query
RO?
Example Set the FX to not create reports.
ROOFF
When you are creating hourly, daily, hourly + daily,
and daily + monthly reports
Syntax
RO p1,p2,p3
p1 Report type
HOUR
Hourly report
DAY
Daily report
HOUR+DAY Hourly and daily reports
DAY+MONTH Daily and monthly reports
p2 Day to create reports (dd; fixed format)
dd
Day (01 to 28)
p3 Hour to create reports (hh; fixed format)
hh
Hour (00 to 23)
Query
RO?
Example Create daily reports at 09:00 every day (in
this case, parameter p2—which is 05 in this
example—is invalid).
RODAY,05,09
Description • You can use this command on models with the
/M1, /PM1, or /PWR1 math option.
• Parameter p2 is invalid if you specify it for
reports other than the monthly and daily
reports.
When you are creating daily + weekly reports
Syntax
RO p1,p2,p3
p1 Report type (DAY+WEEK)
p2 Day of the week to create reports (SUN,
MON, TUE, WED, THU, FRI, SAT)
p3 Hour to create reports (hh; fixed format)
hh
Hour (00 to 23)
Query
RO?
Example Create daily reports at 09:00 every day and
weekly reports at 09:00 every Tuesday.
RODAY+WEEK,TUE,09
Description You can use this command on models with the
/M1, /PM1, or /PWR1 math option.
3.6 Basic Setting Commands