4-32
IM 04L21B01-17EN
Configured Channel Information Data
• The FE5 command is used to output the data.
• The ID number of the output format is 25.
• You can use the CB command to specify whether to output the data of measurement
channels set to skip and computation channels set to OFF.
• The figure below indicates the format.
Number of blocks
Block size
Configured channel information block 1
• • •
Configured channel information block n
(Reserved) (Reserved)
2 bytes
1 byte
1 byte
1 byte
1 byte
Version (Reserved)
2 bytes
Format Details
Item
Description
Output Value
Version
Format version
1
Number of blocks
1
Number of configured channel information blocks
Up to 36
Block size
1
Configured channel information block size
72 (fixed)
Block 1 to n
Configured channel information blocks
Up to 2595 bytes
See Block Details.
1: Output in the byte order specified by the BO command.
Block Details
Each value is the corresponding setting on the FX, but with the decimal point removed.
For example, “120” corresponds to a setting of “1.20” on the FX.
Item
Number
of Bytes
Description
Channel number
1
2
1 to 124
Decimal place
1
0 to 4
When you are using the Log scale,
2
the decimal place of the FX digital
display’s mantissa is displayed.
(Reserved)
1
0
Channel type
1
4
2H for measurement channels. 4H for computation channels.
Logical OR is performed on this value and 800H for channels on which the
range mode is DI, 2400H on channels that are set to Log scale,
2
and 8000H
for channels on which the range mode is skip.
Unit information
8
The terminator is ‘\0.’
Tag information
24
You can enter up to 16 characters for the tag. The terminator is ‘\0.’
Minimum input value
1
4
Measurement channels:
Allowable input range under the current setting
Computation channels:
–9999999, +99999999 (fixed)
Maximum input value
1
4
Span lower limit
1
4
Measurement channels (when scaling is not used): Same value as the FX
span setting
Measurement channels (when scaling is used): Same value as the FX scale
setting
Measurement channels (Log scale
2
): Same value as the FX span setting
Computation channels: Same value as the FX span setting
Span upper limit
1
4
Scale lower limit
1
4
Measurement channels (when the Log scale
2
is not used): The same value
as the span lower or span upper limit value in the above row.
Measurement channels (when the Log scale
2
is used): Same value as the FX
scale setting’s exponent
Computation channels: The same value as the span lower or span upper limit
value in the above row.
Scale upper limit
1
4
FIFO type
1
2
1
Area in the FIFO
1
2
Indicates the position of its own channel in the FIFO block of one sample.The
value starts from zero.
Scale lower limit’s
mantissa
1
2
When the Log scale
2
is not used:
Fixed to 0
When the Log scale
2
is used:
Same value as the FX scale setting
Scale upper limit’s
mantissa
1
2
1
Output in the byte order specified by the BO command.
2
/LG1 option
4.3 Output Format of Binary Data