App index monitoring with a browser – Yokogawa Value Series FX1000 User Manual
Page 37
1-29
IM 04L21B01-17EN
Using the Ethernet Interface (/C7 Option)
1
2
3
4
5
6
App
Index
Monitoring with a Browser
Setting the URL
Set the URL appropriately according to the network environment that you are using. You
can access the FX by setting the URL as follows:
http://host name.domain name/file name
http
Protocol used to access the server.
Host name.domain name
Host name and domain name of the FX.
You can also use the IP address in place of the host name and domain name.
File name
File name of the monitor page and operator page of the FX.
File name of the monitor page: monitor.htm
File name of the operator page: operator.htm
Omitting the file name is equivalent to specifying the monitor page. However, if the
monitor page is disabled, it is equivalent to specifying the operator page.
Example
To display the operator page on a PC in the same domain as the FX, enter the URL in
the Address box of the browser as follows:
http://fx1000.recorder.com/operator.htm or
http://192.168.1.100/operator.htm
(In the example, the domain name is set to recorder.com, the host name to fx1000,
and the IP address to 192.168.1.100.)
Login
You need to configure the following settings to use the login function.
No. Setting
Description and Reference
1
Communication login
(Security > Communication)
To access the FX through a communication interface, you
must log in. For details, see section 8.2 in the FX1000 User’s
Manual, IM 04L21B01-01EN.
2
Login
Register the users who can access the Web server. For
details, see section 8.2 in the FX1000 User’s Manual, IM
04L21B01-01EN.
3
Web page
Set Access control to On in the operator and monitor pages.
Only users whose mode is set to Web, Com, or Key+Com can access the FX Web page.
When you access the page, you will be prompted for a user name and password. Enter
the user name and password that you set in item 2 in the table.
1.5 Monitoring the FX on a PC Browser