beautypg.com

Alarm summary – Yokogawa Value Series FX1000 User Manual

Page 158

background image

4-22

IM 04L21B01-17EN

Alarm Summary

• The FL command is used to output the data.
• The alarm summary is output. Up to 1000 alarm events are retained. Alarm events

that exceed 1000 are cleared from the oldest data.

• Syntax

EACRLF
yy/mo/dd_hh:mm:ss_kkk_ccc_ls_nnnnnnnnnnCRLF
···································
ENCRLF

yy/mo/dd hh:mm:ss

Time when the alarm occurred

yy

Year (

00 to 99)

mo

Month (

01 to 12)

dd

Day (

01 to 31)

hh

Hour (

00 to 23)

mm

Minute (

00 to 59)

ss

Second (

00 to 59)

kkk Alarm cause

OFF:

Alarm release

ON:

Alarm occurrence

ACK:

Alarm acknowledge

ccc Measurement or computation channel number
l Alarm level (1 to 4)
s Alarm type (H, h, L, l, R, r, T, or t)
nnnnnnnnnn Alarm sequence
_ Space

For all-channel alarms, the channel number, alarm level, and alarm status items

are all set to asterisk.

• Example

EA
11/05/11 12:20:00 ON 001 1L

1

11/05/11 12:30:00 OFF 131 3t

2

11/05/11 12:31:00 OFF *** **

3

11/05/11 12:32:00 ACK

4

EN