beautypg.com

3 setup parameters, 3 setup parameters -9, App index – Yokogawa Value Series FX1000 User Manual

Page 91: Example of entering measuring range parameters

background image

3-9

IM 04L21B01-17EN

Commands

1

2

3

4

5

6

App

Index

3.3 Setup Parameters

Example of Entering Measuring Range Parameters

When you enter the span upper and lower limit parameters of the SR command (the
input range setting command), you have to enter all the digits including those to the right
of the decimal point. For example, if you want to set the upper limit to 1.0000 V when the
measuring range is –2.0000 V to 2.0000 V, you have to enter “10000.” If you want to set
the limit to 0.5000 V, you have to enter “5000.”

The table below shows configuration examples.

Measuring

Range

Input Type

Parameter

Selectable Range of the

Measuring Range

Specified Range

Parameter

VOLT

20mV

–20.000mV to 20.000mV

–10.000mV to 20.000mV

–10000 to 20000

/SQRT

2V

–2.0000V to 2.0000V

–2.0000V to 0.5000V

–20000 to 5000

TC

R

0.0 to 1760.0

0.0 to 400.0

0 to 4000

K

–200.0 to 1370.0

–200.0 to 1370.0

–2000 to 13700

RTD

Pt100

–200.0 to 600.0

–10.0 to 500.0

–100 to 5000

DI

LEVEL

0 to 1

0 to 1

0 to 1