App index status information 4, Status information 5 to 8 – Yokogawa Value Series FX1000 User Manual
Page 173
5-3
IM 04L21B01-17EN
Status Reports
1
2
3
4
5
6
App
Index
Status Information 4
Bit
Name
Description
0
A/D conversion complete
Set to 1 when the A/D conversion of the measurement is complete.
1
Medium access complete
Set to 1 when the display, event, manual sampled, report, or screen
image data file are finished being saved to the external storage medium.
Set to 1 when setup data is successfully saved or loaded.
2
Report generation complete
Set to 1 when report generation is complete (/M1, /PM1, and /PWR1
options).
3
Timeout
Set to 1 when the timer expires.
4
–
–
5
–
–
6
USER key detection
Set to 1 when the USER key is pressed.
7
–
–
Status Information 5 to 8
All bits are zeroes.
5.2 Bit Structure of the Status Information
See also other documents in the category Yokogawa Sensors:
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (85 pages)
- EJA110E (4 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJA430E (41 pages)
- EJA430E (96 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)