beautypg.com

App index – Yokogawa Value Series FX1000 User Manual

Page 141

background image

4-5

IM 04L21B01-17EN

Responses

1

2

3

4

5

6

App

Index

if((csum = (csum & 0xffff) + ((csum>>16) & 0xffff)) 0xffff)

/* Add the overflowed digits */

csum = csum - 0xffff;

/* If the digit overflows again, add a 1. */

return((~csum) & 0xffff);

/* bit inversion */

}

Dedicated Commands for RS-422A/485

The dedicated commands for the RS-422A/485 interface and the corresponding
responses are listed in the following table.

Command Syntax Meaning

Response

ESC Oxx CRLF

Opens the device.

• Response from the device with the specified

address

ESC Oxx CRLF

• No response when the device with the specified

address does not exist

1

ESC Cxx CRLF

Closes the instrument. • Response from the device with the specified

address

ESC Cxx CRLF

• No response when the device with the specified

address does not exist

1

1: Some of the possible reasons that cause the condition in which the device with the specified

address cannot be found are a command error, the address not matching that of the device,

the device is not turned ON, and the device not being connected via the serial interface.

• The “

xx” in the table indicates the device address. Specify the address that is

assigned to the instrument from 01 to 99.

• Only one device can be opened at any given time.
• When a device is opened with the ESC O command, all commands on the device

become active.

• When a device is opened with the ESC O command, any other device that is open

is automatically closed.

• Normally, either CR+LF or LF can be used as a terminator for communication

commands. However, the terminator for these commands must be set to CR+LF.

Note

The ASCII code of ESC is 1BH. See appendix 3.

4.1 Response Syntax