App index, Yq sets the communication timeout (/c7 option), Yt sets the ftp transfer timing (/c7 option) – Yokogawa Value Series FX1000 User Manual
Page 123
3-41
IM 04L21B01-17EN
Commands
1
2
3
4
5
6
App
Index
YQ
Sets the communication timeout
(/C7 option)
When you are not using a timeout
Syntax
YQ p1
p1 Communication timeout (OFF)
Query
YQ?
Example Disable the communication timeout.
YQOFF
When you are using a timeout
Syntax
YQ p1,p2
p1 Communication timeout (ON)
p2 Timeout value in minutes (1 to 120)
Query
YQ?
Example Enable the communication timeout, and set the
timeout value to 3 minutes.
YQON,3
YT
Sets the FTP transfer timing (/C7
option)
Syntax
YT p1,p2,p3
p1 Automatically transfer data when display
data files and event data files are created
(ON, OFF)
p2 Automatically transfer data when report data
files are created (ON, OFF)
p3 Automatically transfer data when snapshot
data files are created—when a snapshot is
executed (ON, OFF)
Query
YT?
Example Set the FX to automatically transfer display data
files and event data files. Do not transfer report
data files or screen image data files.
YTON,OFF,OFF
Description • When the method to save data to the external
storage medium is set to “Auto,” the FX
automatically transfers data files when they
are created. For the setting procedure for the
method to save data to the external storage
medium, see the FX1000 User’s Manual, IM
04L21B01-01EN.
• Parameter p2 is only valid on models with the
/M1, /PM1, or /PWR1 math option.
YU
Sets what kind of information to
send using e-mail (/C7 option)
When you want to send changes in the status of
alarms
Syntax
YU p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,
p11,p12
p1 Information to send (ALARM)
p2 Recipient 1 (ON, OFF)
p3 Recipient 2 (ON, OFF)
p4 Transmission of the alarm number 1 status (ON,
OFF)
p5 Transmission of the alarm number 2 status (ON,
OFF)
p6 Transmission of the alarm number 3 status (ON,
OFF)
p7 Transmission of the alarm number 4 status (ON,
OFF)
p8 Inclusion of instantaneous data (ON, OFF)
p9 Inclusion of the source URL (ON, OFF)
p10 Subject (up to 32 characters)
p11 Header 1(up to 64 characters)
p12 Header 2(up to 64 characters)
Query
YU[p1]?
Example Send the status of alarm numbers 1 to 4 to
recipient 1. Include instantaneous data but
not the source URL. Set the subject to “ALM,”
header 1 to “LP2,” and header 2 to “FX.”
YUALARM,ON,OFF,ON,ON,ON,ON,ON,OFF,
ALM,LP2,FX
When you want to send e-mail at scheduled times
Syntax
YU p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,
p11,p12
p1 Information to send (TIME)
p2 Recipient 1 (ON, OFF)
p3 Interval for sending e-mail to recipient 1
(1H, 2H, 3H, 4H, 6H, 8H, 12H, 24H)
p4 Time to send e-mail to recipient 1 (00:00 to
23:59)
p5 Recipient 2 (ON, OFF)
p6 Interval for sending e-mail to recipient 2
(1H, 2H, 3H, 4H, 6H, 8H, 12H, 24H)
p7 Time to send e-mail to recipient 2 (00:00 to
23:59)
p8 Inclusion of instantaneous data (ON, OFF)
p9 Inclusion of the source URL (ON, OFF)
p10 Subject (up to 32 characters)
p11 Header 1(up to 64 characters)
p12 Header 2(up to 64 characters)
Query
YU[p1]?
Example Send an e-mail to recipient 1 every day at 17:15.
Do not include instantaneous data, but include
the source URL. Set the subject to “GOOD” and
header 1 to “LP2.”
YUTIME,ON,24H,17:15,OFF,,,OFF,ON,
GOOD,LP2
3.6 Basic Setting Commands