Operator’s Manual

922244-00 Rev A

35

B

MSRP may assume values ranging from 1 (maximum frequency) up
to 65532 (minimum frequency), by multiples of 4.

The optimal selection of MSRP must take into account both the
maximum frequency of the signal to be generated and its duration.
For example, an ArbStudio 1102/1104 with a maximum frequency of
125 Msps (on the modulating signal) and a buffer depth of 2,096,896
samples in static mode, done by setting SRP = 1 (sampling rate = 125
Msps), you can save a modulating waveform with total length of
16.775168 ms and a time resolution of 8 ns.

Channel Settings - The Run Mode Tab
(Waveform Sequence)

Use the Run Mode tab to define the generation mode for the edited
sequence and set the trigger for the modulating signal.

The Waveform Sequence Run Mode tab allows defining the generation
mode for the edited sequence in Arbitrary mode. The available drop down
Force Trigger options include Single, Continuous, Stepped, and Burst
explained as follows: