Edge AFTER Level - A trigger condition occurs when an edge
condition is detected at least once after one is detected on logic
levels.

ALWAYS - A trigger condition occurs automatically when data
acquisition starts. Any setting on edges or on logic levels is ignored.

NEVER - Trigger is inhibited. Any setting on edges or on logic levels is
ignored. A trigger event, however, can be activated manually during
data acquisition.

Pattern Settings - The Clock Generator Tab

Use the Clock Generator to generate a clock signal with a frequency up to
125 MHz.

The clock is created on Ch15.

It is possible to combine the clock with the Extended Trigger In
vector using the strobe function. The following options are available:
NOP, OR, AND, XOR, NOT, NOR, NAND, and XNOR.

The Clock generator provides a synchronized signal to the Sequencer
Module to generate vectors synchronized with the output clock.