Pulse definition area – Teledyne LeCroy ArbStudio User Manual

Page 149

background image

Operator’s Manual

922244-00 Rev A

141

B

Pulse Definition Area

This section allows defining the pulse for the selected channel.

The following fields are available:

Frequency [Hz] - Set the frequency of the pulse waveform from
100mHZ to 20MHz. Cannel frequencies are independent pairs (1-2 /
3-4 for 1104 ArbStudio instruments). Within the sample pair, the
channel frequency is perfectly matched if one is an even multiple of
the other, otherwise the channel frequency might be rounded to
minimize the frequency % error. The rounded off frequency value is
shown on the Resulting freq. [Hz] field.

Period[s]/Frequency[Hz] - Click the Period(s) button if you want to
define the pulse waveform by its period.

Amplitude[V] - Amplitude of the pulse waveform. It refers to the
impedance load, so if the load is 50 Ω the maximum amplitude is
±6V; if the load is low impedance, the maximum amplitude is ±12V.

Rise Time[%] - Rise time percentage relative to the period of the
pulse waveform.

Fall Time[%] - Fall time percentage relative to the period of the
pulse waveform.

Offset[V] - Voltage offset of the pulse waveform.

Phase[deg] - Phase of the pulse waveform.