beautypg.com

Digital sampler setup example – Teledyne LeCroy ArbStudio User Manual

Page 38

background image

ArbStudio

30

922244-00 Rev A

13. Press the

Run/Stop toolbar button.

NOTE: Once the instrument has started (next step), Single Run Mode
repeats Waveform1 until either the Run/Stop or Force Stop toolbar
buttons are clicked.

The software loads the settings and waveforms into the ArbStudio
hardware and then generates waveforms.

Waveform1 ports to the CH1 BNC output, which can be connected
to an oscilloscope for signal analysis.

Digital Sampler Setup Example

1. Connect an ArbStudio with digital pattern generator capability to

your PC and open the ArbStudio software.

2. Click the

Create Workspace button and follow the steps to

create a new workspace

.

3. Connect the POD probe to Digital POD A.

4. On the Setup wizard, choose Pattern Generator mode.

5. Under POD A, select Signals Definition.

6. Enter CLK in the Bus or Signal Name field and press

to add it to

the list of signals.

7. Select Ch0 from the list of channels on the right and press

to

associate Ch0 with the CLK signal.

8. Repeat steps 6 and 7 to add the WR and RD signals, associating

them with Ch1 and Ch2, respectively.

9. Enter Data in the Bus or Signal Name field and associate it with as

many digital channels as required by the digital data. More than one
channel can be associated with the Data field by selecting all the

required channels and pressing the

button.

10. Double-touch on the Waveform Sequencer of POD A under the

Control Navigation Tree.