B

Rising edge

Falling edge

Any edge

Va (Bus HEX)

Sets a bus at the hexadecimal value entered in the field adjacent (to the
right).

Set the state on a channel by first selecting condition A or B of the desired
level, and then clicking on the corresponding channel row to be set on the
right side listing.

NOTE: The column Signal Names contains the user-defined names for the
signals to acquire.

Set a condition by clicking the desired button on the top. If more than one
signal is selected, the same level or edge setting is applied to the group of
signals.

You can also set a level on all selected signals by writing its value in
hexadecimal format in the HEX field, and then clicking the Va button.

C

ONDITIONS ON EDGES

You can indicate whether the trigger condition must be detected on a
rising, falling, or either edge for each signal.

When trigger conditions are set on edges of several signals, the trigger
event is determined by the occurrence of at least one condition.

If no trigger conditions on edges are set for any signals, trigger events on