Operator’s Manual

922244-00 Rev A

173

B

Click OK to confirm the import of your waveform.

4. On the Project Waveforms tab under the Waveform Sequencer,

select the waveform you just imported and click the Add Waveform
button (adding your newly imported waveform to the Sequencer).

5. Double-click Settings under Channel 1 (Arbitrary) on the Device

Control Tree. (Channel 1 can be any Analog channel you have
configured under Arbitrary Waveform Generator operating mode in
your workspace.)

The DEV1 Settings Ch 1 pop-up screen is shown.

The General Tab is shown by default when the pop-up displays. On
the General Tab be sure to select an Output Impedance value
suitable for your impedance load. For example, if you need to
connect the instrument to an oscilloscope, select 50 Ω as the output
impedance load and 50 Ω as the input impedance on the
oscilloscope channel.