beautypg.com

Introduction – HEIDENHAIN iTNC 530 (340 49x-04) Touch Probe Cycles User Manual

Page 45

background image

HEIDENHAIN iTNC 530

45

2.6 Using the T

o

uc

h Pr

obe F

unctions with

Mec

h

anical Pr

obes or Dial Gaug

es

2.6 Using the Touch Probe

Functions with Mechanical
Probes or Dial Gauges

Introduction

If you do not have an electronic 3-D touch probe on your machine, you
can also use all the previously described manual touch probe functions
(exception: calibration function) with mechanical probes or by simply
touching the workpiece with the tool.

In place of the electronic signal generated automatically by a 3-D touch
probe during probing, you can manually initiate the trigger signal for
capturing the probing position by pressing a key. Proceed as follows:

Select any touch probe function by soft key.

Move the mechanical probe to the first position to be

captured by the TNC.

Confirm the position: Press the actual-position-

capture key for the TNC to save the current position.

Move the mechanical probe to the next position to be

captured by the TNC.

Confirm the position: Press the actual-position-

capture key for the TNC to save the current position.

If required, move to additional positions and capture

as described previously.

Datum:

In the menu window, enter the coordinates of

the new datum, confirm with the SET DATUM soft
key, or write the values to a table (see “Writing the
measured values from touch probe cycles in datum
tables,” page 30, o
r see “Writing the measured
values from touch probe cycles in the preset table,”
page 31).

To terminate the probe function, press the END key.