64

QC-300 Series User’s Guide

Multiple edge probe
The multiple edge probe consists of a standard crosshair with two small concen-

tric circles at the center for positioning the edge detector. Points will be probed

manually for some features as they are when the single edge probe is used, and

automatically for line, circle and arc features when the probe is seeded.

Probe seeding

When points are probed automatically as diagrammed at the right, the probe

must first be seeded by probing the minimum number of points required for

a geometric definition of the feature type. Only lines, circles and arcs can be

probed automatically.

• Line Points probed automatically after seeding with 2 points

• Circle Points probed automatically after seeding with 3 points

• Arc Points probed automatically after seeding with 3 points

• Point Points probed manually - identical to single edge probe

• Distance Points probed manually - identical to single edge probe

• Angle Points probed manually - identical to single edge probe

• Slot Points probed manually - identical to single edge probe

• Rectangle Points probed manually - identical to single edge probe

When probing lines, circles and arcs, the multiple edge probe must be seeded to

provide the system with a general description of the feature location. The location

of seed points is important. The order of seed points is important for arcs.

• Line Seed by probing 2 points at the extreme ends of the line.

• Circle Seed by probing 3 points evenly distributed around the perimeter

of the circle.

• Arc Seed by probing 3 evenly distributed points in sequence from

the beginning to end of the arc.

The seed points can be entered manually by

pressing the Enter key, or automatically using

the auto-enter function.

When all the required points have been seeded, a green arrow and red circle are

displayed. Move the stage to place the green arrow into the red circle to initiate

automatic probing of the remaining points and complete the measurement.

Auto-enter

enabled

Auto-enter

disabled

When the probe is seeded...

move the stage to place...

the arrow in the circle