beautypg.com

Reports are not printed or are incomplete, Reports are printed incorrectly – HEIDENHAIN ND 1300 VED and Crosshair Systems User Manual

Page 314

background image

294

QC-300 Series User’s Guide

Reports are not printed or are incomplete

Probable cause 1: The channel input device is not connected

or is malfunctioning

Possible solution:

1 The printer is not supported by the QC-300.

Probable cause 2: The USB cable is not firmly connected or

is damaged

Possible solution:

1 Connect or replace the printer cable.

Probable cause 3: The printed report setup is incorrect

Possible solution:

1 Correct the report setup in the Print setup screen.

Reports are printed incorrectly

Probable cause 1: The printed report setup is incorrect

Possible solution:

1 Correct the report setup in the Print setup screen.

Probable cause 2: The wrong control characters are specified

Possible solution:

1 Correct the control character configurations in the Print setup screen.