HEIDENHAIN ND 1300 VED and Crosshair Systems User Manual

Page 271

background image

251

11

Setup

4 Exit the setup mode, measure the skew through the

bottom grid points of the calibration grid and create a zero

datum at the lower left point of the grid.

5 Measure the grid values starting at the bottom left corner of

the grid (X1, Y1) and proceed from left to right and from bottom

to top. Be sure that the X and Y values at the lower left corner of

the grid are zero.

6 Delete the skew and zero datum features from the feature

list.

7 Return to the NLEC setup screen.

8 Touch the Calibrate from Features button.

A message will show the resulting machine

zero offset change. If the change seems rea-

sonable, touch OK.

9 Enable NLEC error correction

Zero datum

Skew

NLEC Screen