beautypg.com

HEIDENHAIN ND 1300 VED and Crosshair Systems User Manual

Page 313

background image

293

12

Problem Solving

Probable cause 4: The wrong encoder is connected to the axis

Possible solutions:

1 Connect the axis encoder specified in the Encoders setup screen.

2 Confirm that the each encoder is connected to the correct axis input.

Probable cause 5: The wrong in/mm unit of measure is specified

Possible solution:

1 Specify the correct units of measure in the Encoders setup screen.

Probable cause 6: The wrong encoder reference mark is specified

Possible solution:

1 Correct the reference mark type in the Encoders setup screen.

Probable cause 7: Error correction is required to compensate

encoder inaccuracies:

Possible solutions:

1 Perform LEC, SLEC or optional NLEC error correction in the corresponding error screen.

2 Error correction was performed but not enabled. Enable error correction in the

LEC, SLEC or optional NLEC setup screen.

Probable cause 8: Video resolution is not calibrated:

Possible solutions:

1 Calibrate the video pixel resolution in the VED setup screen.

Probable cause 9: The wrong video camera is specified:

Possible solutions:

1 Specify the correct camera in the VED setup screen.

Symptoms, probable causes & solutions