Status – Measurement Computing Analyzer488 User Manual
Page 170

Section 4
Serial Controller
4.76
STATUS
STATUS
is a system command used to read the Analyzer488 operational status
and data recording conditions. STATUS may be used to return the functional mode,
current IEEE 488 bus address, extended internal status, error status, progress of the
current recording sequence, and information about the most recent record sequence.
This command does not have a query option.
TYPE
System command
SYNTAX
STATUS
[;][number]
number
is 0 to 5. If not specified, 0 is assumed.
RESPONSE
Character string as described below.
REMARKS
The following STATUS information is available:
STATUS
Returns CONTROLLER
Mode, where addr is the IEEE 488 bus device address of the
Analyzer488. Returns the string: "LOW LEVEL", if in Low Level
Controller Mode.
STATUS 1
Returns the string: "LOW LEVEL" if in Low Level Controller
Mode. If in High Level Controller Mode, returns an extended
internal status string in the following format:
STATUS1
string format: m nn Gn s Sn Enn Tn Cn msg
Item Starting Col. # Cols.
Value
m
Operating mode
1
1
C:
Controller
nn
Bus address
3
2
Two-digit decimal number 00 to 30
Gn
Address change
6
2
G0: Address status change has not occurred.
G1: Address status change has occurred.