Dump – Measurement Computing Analyzer488 User Manual
Page 122

Section 4
Serial Controller
4.28
DUMP
The record command, DUMP, returns bus event data from the record memory.
Blocks of data may be dumped by specifying a starting and ending memory location
or a starting location and a length. Issuing DUMP without any options will dump the
entire record memory.
DUMP
returns blocks of data in Motorola S record (S1S9) format, up to 32 data
bytes per line. Any range of locations, from 0 to 32767, may be dumped.
This command does not have a query option.
TYPE
Record command
SYNTAX
DUMP
or
DUMP {location} TO {location}
or
DUMP {location} LENGTH {count}
location is an absolute address of the record memory, from 0 to
32767.
count
is the number of record memory locations to be
dumped
from 0 to 32767.
RESPONSE
Data encoded in S1S9 file format: S1 data records with a maximum
of 32 data bytes per line, followed by an S9 termination record.
REMARKS
S1 data record format:
where (all numbers are in hexadecimal):
type is the record header: S1
record length
is the byte count (two hex characters).
address
is the record memory starting address (four hex
characters). The address is incremented twice for each location
that is dumped. (address is always twice the starting record
memory location.)