Read – Measurement Computing Analyzer488 User Manual
Page 152

Section 4
Serial Controller
4.58
READ
The READ command is used to accept data from the bus when in the Low Level
Controller Mode and return a machine readable bus status string.
READ
may only be used when the Analyzer488 is in the Listen State (see
LISTEN
) and handshaking is set to STEP (see HANDSHAKE). READ performs an
acceptor handshake. The STEP command is not required when using the READ
command since the handshake is automatically performed when READ is executed.
This command does not have a query option.
TYPE
Low Level Controller command
SYNTAX
READ
RESPONSE
Response is dependent on the data currently on the bus. Two
hexadecimal bytes are returned in the following format:
Hex Byte 1
Hex Byte 2
(bit positions)
(bit positions)
(Control and Handshake Lines)
(Data Lines)
8 7 6 5
4 3
2 1
8 7 6 5 4 3 2 1
REN IFC SRQ EOI DAV NRFD NDAC ATN D8
D1
MODE
Low level controller
BUS STATES
None
QUERY
None
RESPONSE
EXAMPLES
PRINT#1,"CONTROLLER OFF"
PRINT#1,"ASSERT IFC ATN REN"
PRINT#1,"UNASSERT IFC"
PRINT#1,"TALK"
PRINT#1,"HANDSHAKE STEP"