- Digital Laser Sensor LV-N10 Series User's Manual -

Limit setting output mode

Setting output 2 to "LiMt" (limit setting output mode) enables detection of reductions in
received light intensity for output 1.
Specifically, if the cumulative minimum of peak value for output 1 falls below the limit
setting value, it is recognized as a reduction in received light intensity and therefore
output 2 is turned on.

"Light intensity hold display" (page 4-24)

Timing chart

For period (1), the peak value
is not below the limit setting
value, so output 2 remains
OFF.
For period (2), the peak value
is below the limit setting value,
so output 2 turns ON.
For period (3), the peak value
again is above the limit setting
value, so output 2 turns OFF if
Auto (Auto) is selected. If USEr
(User) is selected, the
cumulative minimum of peak
values is the peak value
during period (2), which is below the limit setting value, so output 2 remains ON.

"How to reset after the output is activated" (page 4-34)

How to reset after the output is activated

Refer to page 4-3 "2-Output Settings (

)" for setting methods.

OFF

ON

Received light intensity

Peak hold value

Limit setting for output 2

Setting value of received
light intensity for output 1

Output2
(When

#W6Q is selected)

Output2
(When

75'4 is selected)

Output1

OFF

ON

OFF

ON

(2)

(1)

(3)

Display

Resetting method

USEr (User)

User reset

- Press and hold the [MODE] and [SET] buttons simulta-

neously.

- Execute reset using external input.
- Cycle the power.

Auto (Auto)

Auto reset

- Press and hold the [MODE] and [SET] buttons simulta-

neously.

- Execute reset using external input.
- Cycle the power.
- When "peak value > setting value + hysteresis"