Positioning calibration, Positioning calibration -15 – KEYENCE LV-N10 Series User Manual

Page 39

background image

3-3 Adjusting the Sensitivity

3

Ba
sic
Oper

ati
on

3-15

- Digital Laser Sensor LV-N10 Series User's Manual -

This method is used when you want to position a workpiece.

1

Press the [SET] button
once when no workpiece is
present.

2

Position a workpiece such
that its edge aligns with the
center of the projecting
beam. Then, press and hold
the [SET] button for 3 sec-
onds or more. Release the
button when "
SEt" flashes.
Calibration is complete after
the setting value flashes
momentarily, and then stops
(lights up).
Press the

button to

adjust the setting value.

When performing two-point calibration on channel 2 of the 2-output type,
set the channel switch to "2".

Positioning Calibration

Workpiece

Reference

Setting
value

Time

Receiv

ed

lig

ht intensity

2

1

The setting value is determined as the value of light intensity received when
the workpiece comes into position.