TO R Q U E C E L L C A L I B R A T I O N
Torque Cell Calibration
Version 2
WinPEP 7 User Guide
A-5
While installing the calibration weights, you should notice the Torque Gauge on the
Torque Cell Calibration Window moving from 0 to about 500 foot-pounds.
Note: The Torque Gauge may or may not be in this range.
• If the torque cell has been previously calibrated incorrectly or has not been
calibrated for a while, the gauge may show values out of this range until
calibration is complete.
• If you use the bolt pattern in the center of the calibration arm, the gauge will
show values around 250-260 foot-pounds.
Note: Let the torque gauge needle stabilize before clicking Next.
9
From the Span Calibration window (Figure A-3), click Next to continue.
At this point, the value on the gauge should match the value on the calibration
arm.
10 Click Previous to calibrate a different torque cell. Repeat steps 7 through 10.
11 Once you have completed calibrating the torque cell(s), remove the calibration
arm and weights and click Finish.
Figure A-4: Calibration Is Complete Window