Torque cell calibration – Dynojet WinPEP 7 User Manual
Page 110
WinPEP 7 User Guide
A P P E N D I X A
Torque Cell Calibration
A-2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TORQUE CELL CALIBRATION
The Torque Module must be calibrated prior to use. Follow the directions on the
screen exactly. Failure to perform the directions accurately will result in improper
torque values.
1
Verify you are in the MakeRun screen.
2
Verify you are connected to the dyno electronics.
Note: For more information on connecting to the dyno electronics, refer to
“Connecting the Dyno Electronics” on page 3-11.
3
Select Tools
!
MakeRun Options
!
Torque Cell Calibration.
Note: Before proceeding, be sure the eddy current brake is free and clear of any
obstructions. There should not be anything resting on the eddy current brake or
the dynamometer drum during this procedure.
4
Click Next to perform the Zero Calibration.
The Calibration window will appear. The hardware is now zeroing out the torque
cell. If the unit does not calibrate, recheck the setup and retry.
Figure A-1: Zero Calibration Window
Once the Zero Calibration is complete, the Calibration Mass window will appear.
5
Enter the Torque Module calibration value found on the calibration arm.
Note: You must perform this step the first time you calibrate the load cell.
Or