4.4 Methods

792 Basic IC

89

European Pharmacopoeia

Formula setting according to European Phar-

macopoeia.

US Pharmacopoeia

Formula setting according to US Pharmaco-

poeia.

Void time/volume

Calculation of

void time/void volume:

Calculation method

Selection of the method for void time calcula-

tion:

None

Manual entry in the

Void time

field.

1st

component

The peak corresponding to the first component

is selected as a void time marker and its reten-

tion time replaces the previous value of the void

time. If the first component is not identified, the

expected retention time is used.

1st peak

The first detected peak is used as a void time

marker for the run and its retention time is

stored in the

Void time

field.

From void volume %

For the calculation of the void time, the % value

entered in the

Void volume

field is multiplied by

the ratio of empty column volume and eluent

flow rate.

Void volume

Void volume in mL and % of column volume.

The % value can be entered manually if

Calcula-

tion method =

From void volume %

is set.

Void time

Void time in s. This value can be entered

manually if

Calculation method =

None

is set.

Index

Calculation of

Retention index:

Interpolation

Use of linear or logarithmic interpolation scale

for retention index calculation.

Internal

The index scale is constructed on the basis of

the current chromatogram. All components

used for index scale calibration should be pre-

sent in the current sample.

External

The index scale is constructed based on an-

other standard chromatogram.