3 system functions – Metrohm 792 Basic IC User Manual
Page 78
4.3 Systems
792 Basic IC
69
4.3.3 System
functions
Start/stop hardware and record baseline
SYSTEM / Control / Startup hardware (Measure Baseline)
Starting the hardware at the 792 Basic IC includes sending of
System
startup values
, starting of the high-pressure pump and starting of the
peristaltic pump.
At the same time, the
recording of the measurement signal using
the method defined for the system is started. Independently of the set
chromatogram
Duration
, the measurement signal is recorded until a new
determination is started. Alternatively the baseline recording can be
stopped by clicking the icon of the chromatogram window. In this
case, the user is asked if the recorded baseline should be saved or not.
SYSTEM / Control / Shutdown hardware
High-pressure pump and peristaltic pump at the 792 Basic IC are im-
mediately stopped. A running determination is also stopped.
Start/stop determinations
SYSTEM / Control / Start determination
Start determination using the settings of the selected system. At this
start command, the
System startup values
are set at the 792 Basic IC.
The high-pressure pump and the peristaltic pump are started if they are
not already running. The time program is started immediately; the data
recording is started after switching the injection valve to the "Inject"
position.
At each start of a determination, the
Edit sample description
window is
opened automatically. This window can be used for entering informa-
tion concerning the sample: