6.1 Technical data

792 Basic IC

161

6.1.5 Peristaltic

pump

Type

2-channel peristaltic pump

Pump capacity

Rotational speed

20 U/min at 50 Hz

24 U/min at 60 Hz

Flow range

0.5}0.6 mL/min with 6.1826.060 Pump tubing

Maximum error

r 5 %

Maximum pressure

0.4 MPa (4 bar)

Pumpable liquids

Clear liquids with no solid contents

Pump tubing material

PP (polypropylene)

6.1.6 Suppressor

module

Switching duration

140 ms

Pressure resistance

2.5 MPa (25 bar)

6.1.7 RS232

interface

Connector

Dsub 9 pin (male)

Function

TxD and RxD signal for connection with software

handshake

Default settings

9600 baud, 8 bit, 1 stop bit, no parity, XON/XOFF

Pin assignment