4.6 Batch reprocessing
792 Basic IC
137
4.6.3
Batch reprocessing queue editor
Open batch reprocessing queue editor window
REPROCESS /
This menu item opens the batch reprocessing editor program for edit-
ing the batch reprocessing table.
No
Row
number.
File name
Name of the chromatogram file (read-only).
Method
Name of the method file
*.mtw
used for re-
cording the chromatogram (read-only).
Ident
User defined identifier for chromatogram. Will
be placed into appropriate passport field when
starting the chromatogram.
Vial
Autosampler vial position to take sample from.
Will be placed into appropriate passport field.
Volume
Injected volume of sample in PL. Will be placed
into appropriate passport field.
Dilution
Sample dilution prior to injection. Will be placed
into appropriate passport field.
Amount
Sample amount. If this value is different for the
calibration run (c) and the sample run (s), the
component concentrations of the sample are
calculated as follows:
C
s
=
C
c
•
Amount
s
/
Amount
c
Will be placed into appropriate passport field.
Internal standard amount
Concentration of the internal standard compo-
nent for relative concentration calculations. Will
be placed into appropriate passport field.
Calibration level
Calibration level (see section 4.4.5) for the sam-
ple: Level
0
stays for normal analysis, levels
1
and greater for calibration runs. Correct filling of
this column enables to calculate calibration
coefficients automatically.