Metrohm 792 Basic IC User Manual
Page 125
4 Operation
792 Basic IC
116
raw concentration
Absolute concentrations (raw quantity) of the
components, calculated as
C
i
= W
i
(R
i
) / V'
Total value for the column is the sum of con-
centrations for all components.
concentration%
Normalization of concentrations for all peaks
using the NORM value entered in the
Total % for
normalization
field (default setting 100%):
C
i
% = NORM
•
W
i
(R
i
) /
W
i
(R
i
)
Total value for the column equals NORM for
the main table, and the sum of concentrations
in the group for the group table.
rel. concentration
Relative concentration of the component rela-
tive to the standard component, assumes con-
centration of one of the components to be
known in advance. The concentration of the
component is calculated by the formula:
C'
i
= W
i
(R
i
) / V' = W
i
(R
i
)
•
C
s
/ W
s
(R
s
)
,
where V' = W
s
(R
s
) / C
s
is the effective (redu-
ced) volume of injected sample. The standard
component concentration C
s
is entered by the
user.
Total value for relative concentration column
does not include the concentration of the stan-
dard component.
rel. concentration%
Relative concentration percent differs from
con-
centration%
by exclusion of standard compo-
nent(s) from summation, so that sum of all con-
centrations (excluding standard) equals
NORM
.
Total value for the column equals NORM value
for main peak table, and the sum of concentra-
tions in the group for Group table.
index
Linear or logarithmic retention index for identi-
fied components.
Total value for the column is a weighted index
average with absolute concentration as weight.
I =
(I
i
C
i
) /
C
i
.