3 injection valve, 4 high-pressure pump – Metrohm 792 Basic IC User Manual
Page 169
6 Appendix
792 Basic IC
160
6.1.3 Injection
valve
Actuator switching duration
100}150 ms
Pressure resistance
25 MPa (250 bar)
6.1.4 High-pressure
pump
Type
Serial dual piston pump with two valves
Pump capacity
Flow range
0.20...2.5 mL/min
Maximum error
< r 2 % of set value
Flow constancy
< 0.5 % of set value
Reproducibility of eluent
flow
typ. better than r 0.1 %
Pressure measurement
Pressure range
0...25.0 MPa (0...250 bar)
Residual pulsation
< 10 % (at 1 mL/min water and 10 MPa pressure,
without pulsation dampener)
Measurement principle
Piezoresistive measurement
Response time: 3 ms
Measurement volume: ca. 50 PL
Maximum error
r 3 % of set value
Resolution
0.1 MPa (conductivity measurements)
Sampling rate
1 measurement/piston stroke (pump running)
1 measurement/s (pump not running)
Safety shutdown
Function
Automatic shutdown when upper and lower
pressure limits violated
Maximum pressure limit
adjustable between 0.1...25.0 MPa (1...250 bar)
Response time: 1 pump cycle
Minimum pressure limit
adjustable between 0.1 ... 25.0 MPa (1...250 bar),
inactive at 0 MPa
Response time: 5 pump cycles
Pump head
Pump head volumes Main
piston: 40
PL
Priming piston:
20 PL
Pump displacement
volumes
Main piston:
28.5 PL
Priming piston:
14.25 PL
Length of stroke
Main piston:
3.6 mm
Priming piston:
1.8 mm