beautypg.com

3 injection valve, 4 high-pressure pump – Metrohm 792 Basic IC User Manual

Page 169

background image

6 Appendix

792 Basic IC

160

6.1.3 Injection

valve

Actuator switching duration

100}150 ms

Pressure resistance

25 MPa (250 bar)



6.1.4 High-pressure

pump

Type

Serial dual piston pump with two valves

Pump capacity

Flow range

0.20...2.5 mL/min

Maximum error

< r 2 % of set value

Flow constancy

< 0.5 % of set value

Reproducibility of eluent

flow

typ. better than r 0.1 %

Pressure measurement

Pressure range

0...25.0 MPa (0...250 bar)

Residual pulsation

< 10 % (at 1 mL/min water and 10 MPa pressure,

without pulsation dampener)

Measurement principle

Piezoresistive measurement

Response time: 3 ms

Measurement volume: ca. 50 PL

Maximum error

r 3 % of set value

Resolution

0.1 MPa (conductivity measurements)

Sampling rate

1 measurement/piston stroke (pump running)

1 measurement/s (pump not running)

Safety shutdown

Function

Automatic shutdown when upper and lower

pressure limits violated

Maximum pressure limit

adjustable between 0.1...25.0 MPa (1...250 bar)

Response time: 1 pump cycle

Minimum pressure limit

adjustable between 0.1 ... 25.0 MPa (1...250 bar),

inactive at 0 MPa

Response time: 5 pump cycles

Pump head

Pump head volumes Main

piston: 40

PL

Priming piston:

20 PL

Pump displacement

volumes

Main piston:

28.5 PL

Priming piston:

14.25 PL

Length of stroke

Main piston:

3.6 mm

Priming piston:

1.8 mm