3 connection of supply bottles – Metrohm 792 Basic IC User Manual
Page 47
2 Installation
792 Basic IC
38
39
39
39
39 Suppressor module
74
74
74
74
Coupling (6.2744.110)
40
40
40
40 Tubing cartridge (6.2755.000)
for pump tubings 71/72
71/72
71/72
71/72
75
75
75
75 Suppressor inlet capillary for
eluent
41
41
41
41 Contact pressure lever
for adjusting the contact pressure
76
76
76
76 Suppressor outlet capillary for
eluent
42
42
42
42 Holding clamp
for locking the tubing cartridge into
place
77
77
77
77 Suppressor inlet capillary for
H
2
O
43
43
43
43 Snap-action lever
for releasing the tubing cartridge
78
78
78
78 Suppressor inlet capillary for
H
2
SO
4
44
44
44
44 Pump drive
Roller head with contact rollers
79
79
79
79 Suppressor outlet capillary for H
2
O
45
45
45
45 Mounting pin
for attaching the tubing cartridge
80
80
80
80 Suppressor outlet capillary for
H
2
SO
4
2.8.3
Connection of supply bottles
The supply lines for the regeneration and rinsing solution between the
storage bottles and the peristaltic pump are installed as follows:
1 Prepare supply bottle for H
2
SO
4
x Prepare regeneration solution suited for the desired applica-
tion and separating column (normally 20 mmol/L H
2
SO
4
).
x Fill regeneration solution into supply bottle and label the
bottle.
2 Connect aspirating tubing for H
2
SO
4
x Prepare aspirating tubing 69
69
69
69: Cut a piece of the 6.1803.020
PTFE tubing to the required length (normally ca. 120 cm).
x Insert one end of the aspirating tubing 69
69
69
69 into the regenera-
tion solution storage bottle and tighten it so that the tubing is
firmly held.
x Insert the free end of aspirating tubing 69
69
69
69 into one of the
openings 5
555 of the 792 Basic IC (see Fig. 2) and pull it suffi-
ciently far into the interior.
x Mount a coupling 70
70
70
70 (6.2744.030) to the inlet end of the rear
pump tubing 71
71
71
71.
x Mount a 6.2744.010 Compression fitting at the end of the
aspirating tubing 69
69
69
69 and screw this compression fitting on to
the coupling 70
70
70
70 (see Fig. 14).
3 Prepare supply bottle for H
2
O
x Prepare rinsing solution suited for the desired application and
separating column (normally dist. H
2
O).
x Fill rinsing solution into a supply bottle and label the bottle.