2 Installation

792 Basic IC

38

39

39

39

39 Suppressor module

74

74

74

74

Coupling (6.2744.110)

40

40

40

40 Tubing cartridge (6.2755.000)

for pump tubings 71/72

71/72

71/72

71/72

75

75

75

75 Suppressor inlet capillary for

eluent

41

41

41

41 Contact pressure lever

for adjusting the contact pressure

76

76

76

76 Suppressor outlet capillary for

eluent

42

42

42

42 Holding clamp

for locking the tubing cartridge into

place

77

77

77

77 Suppressor inlet capillary for

H

2

O

43

43

43

43 Snap-action lever

for releasing the tubing cartridge

78

78

78

78 Suppressor inlet capillary for

H

2

SO

4

44

44

44

44 Pump drive

Roller head with contact rollers

79

79

79

79 Suppressor outlet capillary for H

2

O

45

45

45

45 Mounting pin

for attaching the tubing cartridge

80

80

80

80 Suppressor outlet capillary for

H

2

SO

4

2.8.3

Connection of supply bottles

The supply lines for the regeneration and rinsing solution between the

storage bottles and the peristaltic pump are installed as follows:

1 Prepare supply bottle for H

2

SO

4

x Prepare regeneration solution suited for the desired applica-

tion and separating column (normally 20 mmol/L H

2

SO

4

).

x Fill regeneration solution into supply bottle and label the

bottle.

2 Connect aspirating tubing for H

2

SO

4

x Prepare aspirating tubing 69

69

69

69: Cut a piece of the 6.1803.020

PTFE tubing to the required length (normally ca. 120 cm).

x Insert one end of the aspirating tubing 69

69

69

69 into the regenera-

tion solution storage bottle and tighten it so that the tubing is

firmly held.

x Insert the free end of aspirating tubing 69

69

69

69 into one of the

openings 5

555 of the 792 Basic IC (see Fig. 2) and pull it suffi-

ciently far into the interior.

x Mount a coupling 70

70

70

70 (6.2744.030) to the inlet end of the rear

pump tubing 71

71

71

71.

x Mount a 6.2744.010 Compression fitting at the end of the

aspirating tubing 69

69

69

69 and screw this compression fitting on to

the coupling 70

70

70

70 (see Fig. 14).

3 Prepare supply bottle for H

2

O

x Prepare rinsing solution suited for the desired application and

separating column (normally dist. H

2

O).

x Fill rinsing solution into a supply bottle and label the bottle.