beautypg.com

Metrohm 792 Basic IC User Manual

Page 88

background image

4.3 Systems

792 Basic IC

79

Error messages

Detection of hard-
ware failed

Bad connection between PC and 792 Basic IC or
instrument switched off (check connecting cable or
switch on instrument).

E1

Program checksum wrong (call Metrohm service).

E2

RAM faulty (call Metrohm service).

E200

Invalid instrument adjustment (call Metrohm service).

E237

Storage of configuration values failed (repeat last
action; if error reappears, call Metrohm service).

E238

Storage of instrument number failed (repeat last
action; if error reappears, call Metrohm service).

E240

EEPROM faulty (call Metrohm service).

E258

Storage of setup values failed (repeat last action; if
error reappears, call Metrohm service).

E295

Storage of memory values failed (repeat last action; if
error reappears, call Metrohm service).

E296

Instrument stopped (restart instrument; if error
reappears, call Metrohm service).

E298

Storage of flow correction value failed (repeat last
action; if error reappears, call Metrohm service).

E299

Storage of break time values failed (repeat last action;
if error reappears, call Metrohm service).

E300

High-pressure pump faulty (restart pump; if error
reappears, call Metrohm service).

E301

Injection valve blocked (check injection valve; if error
reappears, call Metrohm service).

E302

Suppressor module blocked (check suppressor; if
error reappears, call Metrohm service).

E303

Storage of maintenance information failed (repeat last
action; if error reappears, call Metrohm service).