4 Operation
792 Basic IC
98
Additionally, the
Components table
contains the following buttons:
Add a new component (an empty column) to
the component table.
Delete the current component from the compo-
nent table.
Open
the
Peak identification
window (see Identi-
fication).
Open
the
Concentration table
(see Concentration
table).
Open the component window with display of
calibration curve and calibration parameters.
Accept all changes and close the
Component ta-
ble
.
Reject all changes and close the
Component ta-
ble
.
The standard method
Default.mtw
supplied with the «792 Basic IC» pro-
gram has an empty component table. All other methods contain a
component table with pre-defined components with entries for the
names and retention times.
If the output of all peaks is required in the result report then a so-
called universal component can be defined in the component table
for this purpose; this receives the value
0
in the
Time
field and no peak
is allocated to it. The name entered under
Name
(e.g.
unknown
) is then
used for all peaks for which no component can be allocated.
Peak identification
792 BASIC IC / Method / Calibration / Identification
Selecting this menu item or clicking on
in the component
table opens the
Peak identification
window with parameters for tuning the
peak identification procedure.
Number of components
Number of components in the component table
(read-only).
Scheme
Quick choice of identification parameters
scheme.
Standard
Set default meanings of identification parame-
ters:
Height
for reference components,
Time
for
all other components (ordinary components).
Nonstandard
Set custom meanings of identification parame-
ters.