2.7.8

Installation of the separating column with suppressor

For operation of the 792 Basic IC with suppressor module, the IC sepa-

rating column is first connected to the injector or precolumn. The con-

nection to the suppressor module and the detector block is described

in section 2.8.

1 Connect column to injector

x Remove end caps from column 66

66

66

66.

x Screw inlet end of separating column 66

66

66

66 (note flow direction)

to column connection capillary 20

20

20

20 or to the already installed

precolumn (procedure see section 2.7.7).

2 Rinse column

x Place a beaker beneath the column outlet.
x Open software window for manual system control.
x If necessary, modify

Flow rate

to the value suited for the

inserted separating column and click on

to

send this value to the 792 Basic IC.

x Switch on high-pressure pump (

IC pump

) by clicking

and

rinse column with eluent for ca. 10 min.

x Switch off high-pressure pump by clicking

..

3 Fix column

x Insert one or two column holders 67

67

67

67 (6.2027.030, 6.2027.040

or 6.2027.050) in the mounting rails 19

19

19

19 and fasten separating

column 66

66

66

66 in the column holder 67

67

67

67.