beautypg.com

Metrohm 792 Basic IC User Manual

Page 101

background image

4 Operation

792 Basic IC

92

Min height

Minimum height of peaks to be detected.

Negative peaks

If this option is enabled, negative peaks are

also detected.


Events

Tab

Events

of

Integration parameters

window for definition of integration

events. Integration events have a higher priority than integration pa-

rameters and are used for fine-tuning of the integration process. They

should be used only in the case when problems cannot be solved by

tuning a set of parameters from the integration parameters window.

Number of events

Number of integration events (read-only).

Add event to the list.

Edit the selected event.

Delete the selected event.

Accept integration parameters and perform re-

integration.

If an integration event is added or modified, the

Edit integration event

window is opened where the following parameters can be set: