4.4 Methods

792 Basic IC

113

component the summary for all unidentified

peaks is presented as a single line.

Calls up a box with a list of columns that can be

included into the report (see Report elements).

It is available only if

Quantification method = Cus-

tom

has been selected.

Report all peaks

Checkbox that includes all peaks irrespective of

the concentration of the substance. Otherwise

the report generator excludes lines with zero

concentration of the component. It is available

only if

Quantification method = Custom

has been

selected.

Groups

If this checkbox is checked, a separate peak

table report is put out for every group specified

in the component table.

Template options
For report output using a report template the following parameters can

be modified:

Template

Selection of the report template file

*.rtt

. Several

templates for different languages and applica-

tions are available.

Separator

Defines which character separates report col-

umns in tables. A separator other than

Space

is

useful when the report is printed to a file and

this file is imported into a spreadsheet software.

Tab size

Sets tab stop value for the screen window. Im-

portant only when

Separator = Tabulation

has

been selected or there are tab characters in the

template file

*.rtt

.

File output options
If

File

has been selected for the report destination, the following pa-

rameters can be modified:

Directory

Directory for report output. For creation of a

new directory, use the

button.

Name

File name to save report to. The report is saved

in text form in the ANSI or ASCII format. So add

an extension like

*.txt

to the file name. If the

chromatogram plot checkbox is checked, the

plot is saved in a separate file

*.wmf

in the WMF

format under the same name.

Mode

The two options

Overwrite

(overwrite the file) or

Append

(append to existing file) are available.