4 Operation

792 Basic IC

74

Valve

Injection

valve

Switch injection valve to "INJECT" position.

Switch injection valve to "FILL" position

Suppressor

Suppressor

module

Switch the suppressor module to the next posi-

tion. The time since the last switching of the

suppressor module is displayed in the field be-

side the

button.

IC pump

High-pressure

pump

Start high-pressure pump.

Stop high-pressure pump.

Peristaltic

Peristaltic

pump

Start peristaltic pump.

Stop peristaltic pump.

System startup values

The system startup values are sent and applied

to the 792 Basic IC each time the system is

connected, a determination is started, or the

values are sent manually with

.

Flow, mL/min

Startup value for flow rate of the high-pressure

pump.

Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

25.0 MPa

Pressure min, MPa

Startup value for minimum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

25.0 MPa

Time program

On the

Program

tab of the system settings window a user-defined time

program for instrument control can be entered. This program is started

automatically at the moment the determination is started. The time pro-

gram contains program steps including time, program instruction and

parameter.