2.7 Precolumns and separating columns

792 Basic IC

31

64

64

64

64

62

62

62

62

65

65

65

65

46

46

46

46

20

20

20

20

63

63

63

63

62

62

62

62

Fig. 11: Installation of precolumn glass cartridges with cartridge holder

20

20

20

20 Column connection capillary

from injector

63

63

63

63

Sleeve for precolumn cartridge

46

46

46

46

Compression fitting (6.2744.010)

64

64

64

64

Precolumn cartridge (6.1006.030)

48

48

48

48

PEEK capillary (6.1831.010)

65

65

65

65

Connection piece

for connection precolumn – column

62

62

62

62

End fitting

2.7.4

IC anion precolumn SUPERSEP

The 6.1009.010 IC Anion Precolumn SUPERSEP has two connections

for PEEK capillaries and is installed as follows:

1 Connect precolumn

x Remove end caps from the precolumn.
x Fit PEEK compression fitting 46

46

46

46 to column connection capil-

lary 20

20

20

20 (see section 2.3.4).

x Screw precolumn to column connection capillary 20

20

20

20.

x Cut a piece from the 6.1831.010 PEEK capillary as short as

possible and fit with PEEK compression fittings 46

46

46

46 (see sec-

tion 2.3.4).

x Fasten the prepared capillary to the other end of the precol-

umn.

2 Rinse precolumn

x Place a beaker beneath the outlet capillary of the precolumn.
x Open software window for manual system control.
x If necessary, modify

Flow rate

to the value suited for the

inserted separating column and click on

to

send this value to the 792 Basic IC.

x Switch on high-pressure pump (

IC pump

) by clicking

and

rinse precolumn with eluent for ca. 10 min.

x Switch off high-pressure pump by clicking

.