3.3 Calibration

792 Basic IC

49

9 Switch injection valve to "FILL" position

Ÿ Click on the

button of the 792 icon in the system

window to switch the injection valve to the "FILL" position. At

the same time, the suppressor module is switched to the next

position.

10 Fill sample loop

Ÿ Immerse the aspirating tubing 23

23

23

23 in the standard solution.

Ÿ Using the syringe fixed to syringe tubing 25

25

25

25, siphon in ca.

1 mL standard solution.

11 Switch injection valve to "INJECT" position

Ÿ Click on the

button of the 792 icon in the system

window to switch the injection valve to the "INJECT" position.

At the same time, the data recording is started automatically.

The status bar of the chromatogram window shows the mes-

sage

Measure

, the

SYSTEM STATE

window shows the message

INJECT done [792COM#]

.

Ÿ When the analysis time of 16 min set in the method has

expired, the recorded chromatogram is automatically evalu-

ated and stored. The chromatogram window then closes.

12 Open chromatogram

Ÿ Click on

or

File / Open / Chromatogram

in the main win-

dow. Select the chromatogram file

*.chw

just recorded and

click on

. The chromatogram window is opened where

the found peaks are numbered and the baselines are drawn

in.