4.5 Chromatograms

792 Basic IC

131

Subtraction of a chromatogram

792 BASIC IC / Process / More / Subtract

This menu item opens the

Subtract chromatogram

window allowing the

subtraction of any opened chromatogram from the active, selected

chromatogram.

Select the chromatogram that should be subtracted, click on

and then on

792 BASIC IC / View / View all

. The result is shown in the ac-

tive chromatogram window, which is considered as a new chroma-

togram and will be stored under a new name to avoid overwriting of old

data.

Data compression

792 BASIC IC / Process / More / Compress

Selection of this menu item opens the

Raw data compression

window for

compression of the raw data of the active chromatogram by summing

up several neighboring points.

Divisor

Coefficient of compression (the number of data points is

reduced by this value). The default value of the compres-

sion coefficient is calculated so that the width at half-height

of the narrowest peak will be digitized by at least 30 points.

If the user enters a coefficient of compression that exceeds

the value offered by the software, integration precision may

decrease.

Invert chromatogram

792 BASIC IC / Process / More / Invert!

This menu item inverts the response curves for all channels of the

chromatogram so that negative peaks become positive and vice versa

(useful for chromatograms with wrong input polarity).