Metrohm 792 Basic IC User Manual
Page 140
4.5 Chromatograms
792 Basic IC
131
Subtraction of a chromatogram
792 BASIC IC / Process / More / Subtract
This menu item opens the
Subtract chromatogram
window allowing the
subtraction of any opened chromatogram from the active, selected
chromatogram.
Select the chromatogram that should be subtracted, click on
and then on
792 BASIC IC / View / View all
. The result is shown in the ac-
tive chromatogram window, which is considered as a new chroma-
togram and will be stored under a new name to avoid overwriting of old
data.
Data compression
792 BASIC IC / Process / More / Compress
Selection of this menu item opens the
Raw data compression
window for
compression of the raw data of the active chromatogram by summing
up several neighboring points.
Divisor
Coefficient of compression (the number of data points is
reduced by this value). The default value of the compres-
sion coefficient is calculated so that the width at half-height
of the narrowest peak will be digitized by at least 30 points.
If the user enters a coefficient of compression that exceeds
the value offered by the software, integration precision may
decrease.
Invert chromatogram
792 BASIC IC / Process / More / Invert!
This menu item inverts the response curves for all channels of the
chromatogram so that negative peaks become positive and vice versa
(useful for chromatograms with wrong input polarity).