2 Installation

792 Basic IC

36

x Press contact pressure lever 41

41

41

41 on both tubing cartridges

down as far as it will go.

x Insert the pump tubings 71

71

71

71 and 72

72

72

72 (6.1826.060) into each of

the tubing cartridges as shown in Fig. 13. The white-yellow

stopper 73

73

73

73 must click into the corresponding holder on the

left-hand side of the tubing cartridge.

x Place the tubing cartridges on mounting pin 45

45

45

45 and press

down on the right-hand side until snap-action lever 43

43

43

43 clicks

into position on holding clamp 42

42

42

42. Take care that no kinks are

formed in the pump tubing.

2 Install Filter units PEEK

x Mount a coupling 74

74

74

74 (6.2744.110) to the outlet end of the two

pump tubings 71

71

71

71 and 72

72

72

72.

x Attach a filter unit PEEK 28

28

28

28 (see section 2.3.5) to the other

end of this coupling (note flow direction!).

68/69

68/69

68/69

68/69

70

70

70

70 71/72

71/72

71/72

71/72 73

73

73

73

40

40

40

40

41

41

41

41

73

73

73

73

74

74

74

74

77/78

77/78

77/78

77/78

43

43

43

43

46

46

46

46

28

28

28

28

46

46

46

46

Fig. 13: Installing pump tubings

28

28

28

28

Filter unit PEEK (6.2821.100)

70

70

70

70 Coupling (6.2744.030)

40

40

40

40

Tubing cartridge

71

71

71

71 Pump tubing (6.1826.060) for H

2

SO

4

41

41

41

41

Contact pressure lever

72

72

72

72 Pump tubing (6.1826.060) for H

2

O

43

43

43

43 Snap-action lever

73

73

73

73 Stopper (white-yellow)

46

46

46

46

PEEK compression fitting

(6.2744.010)

74

74

74

74 Coupling (6.2744.110)

68

68

68

68 Aspirating tubing for H

2

O

77

77

77

77 Suppressor inlet capillary for H

2

O

69

69

69

69 Aspirating tubing for H

2

SO

4

78

78

78

78 Suppressor inlet capillary for

H

2

SO

4