3 operating tutorial, 1 requirements – Metrohm 792 Basic IC User Manual
Page 54
3.1 Requirements
792 Basic IC
45
3 Operating
tutorial
This section introduces you to the operation of the 792 Basic IC by
means of a brief operating tutorial which describes the basic operat-
ing steps needed for the recording of an ion chromatogram using one
of the system files supplied.
The determination of the anionic content of a drinking water sample
with the METROSEP Anion Dual 1 column is used as an illustrative
example. Please note that the steps and parameter settings described
apply only to this example. If you use a different separating column
and system file, the procedures described in the tutorial must be
modified appropriately.
For further explanations of the operation, please refer to section 4.
3.1 Requirements
For the determination of anions in drinking water described in this tuto-
rial, the following instruments, accessories and solutions are required:
x 2.792.0020 Basic IC
with suppressor module
x 6.1825.210 Sample loop (20 P
P
P
PL, PEEK)
already integrated in the 792 Basic IC
x 6.5325.000 Starter Kit
incl. the following accessories:
6.1006.020 IC Column cartridge
METROSEP Anion Dual 1 (150 mm)
6.2828.100 Glass cartridge holder (150 mm)
6.2620.150 Pulsation dampener
x Eluent
2.4 mmol/L NaHCO
3
/ 2.5 mmol/L Na
2
CO
3
/ 2 % Acetone
in dist. H
2
O
Flow: 0.5 mL/min
x Standard
Standard solution with 0.5 mg/L fluoride, 5 mg/L chloride, 5 mg/L
nitrite, 10 mg/L bromide, 10 mg/L nitrate, 10 mg/L phosphate and
10 mg/L sulfate (in dist. H
2
O)