Page 54

background image

3.1 Requirements

792 Basic IC

45

3 Operating

tutorial


This section introduces you to the operation of the 792 Basic IC by

means of a brief operating tutorial which describes the basic operat-

ing steps needed for the recording of an ion chromatogram using one

of the system files supplied.
The determination of the anionic content of a drinking water sample

with the METROSEP Anion Dual 1 column is used as an illustrative

example. Please note that the steps and parameter settings described

apply only to this example. If you use a different separating column

and system file, the procedures described in the tutorial must be

modified appropriately.
For further explanations of the operation, please refer to section 4.





3.1 Requirements

For the determination of anions in drinking water described in this tuto-

rial, the following instruments, accessories and solutions are required:

x 2.792.0020 Basic IC

with suppressor module

x 6.1825.210 Sample loop (20 P

P

P

PL, PEEK)

already integrated in the 792 Basic IC

x 6.5325.000 Starter Kit

incl. the following accessories:

6.1006.020 IC Column cartridge

METROSEP Anion Dual 1 (150 mm)

6.2828.100 Glass cartridge holder (150 mm)
6.2620.150 Pulsation dampener

x Eluent

2.4 mmol/L NaHCO

3

/ 2.5 mmol/L Na

2

CO

3

/ 2 % Acetone

in dist. H

2

O

Flow: 0.5 mL/min

x Standard

Standard solution with 0.5 mg/L fluoride, 5 mg/L chloride, 5 mg/L

nitrite, 10 mg/L bromide, 10 mg/L nitrate, 10 mg/L phosphate and

10 mg/L sulfate (in dist. H

2

O)