2 Installation

792 Basic IC

28

2.7

Precolumns and separating columns

2.7.1

General information on precolumns

The use of easily exchangeable precolumns protects the separating

columns and prolongs their lifetime. The precolumns available from

Metrohm (see section 6.3.2) are either real precolumns or precolumn

cartridges, which are used together with the 6.2821.050 Twin cartridge

holder or the 6.2828.110 Precolumn cartridge holder.

New IC precolumns are normally filled with solution and sealed at both

ends. Before the precolumn is installed in the system, it must be

ensured that this solution is freely miscible with the eluent used

(check manufacturer's specifications).

2.7.2

Precolumns with twin cartridge holder

The 6.1005.020, 6.1005.040, 6.1005.050, 6.1007.010 and 6.1010.010

Precolumn cartridges are installed in the 6.2821.050 Twin cartridge

holder as follows (see Fig. 10):

1 Install cartridge

x Insert inlet capillary 55

55

55

55 with end piece for precolumn cartridge

in Manufit housing 56

56

56

56.

x Insert outlet capillary 61

61

61

61 with end piece for precolumn car-

tridge in Manufit pressure screw 60

60

60

60.

x Remove end caps from the precolumn cartridge 59

59

59

59 (the steel

mesh 58

58

58

58 and gaskets 57

57

57

57 are already inserted in the car-

tridge).

x Fit the two capillary end pieces onto the precolumn cartridge

59

59

59

59 (comply with the flow direction if specified on the column).

x Screw Manufit pressure screw 60

60

60

60 and Manufit housing 56

56

56

56

firmly together.

2 Connect precolumn

x Cut inlet capillary 55

55

55

55 of the assembled precolumn to the

desired length and mount a compression fitting 46

46

46

46 (see sec-

tion 2.5).

x Connect inlet capillary 55

55

55

55 either using the Coupling 26

26

26

26 to the

column connection capillary 20

20

20

20 mounted on the injection

valve or directly to connection "4“ of injection valve A.

x Shorten outlet capillary 61

61

61

61 of the precolumn to ca. 5 cm and

mount a compression fitting 46

46

46

46 (see section 2.5).