Metrohm 792 Basic IC User Manual
Page 154
5.2 Maintenance and servicing
792 Basic IC
145
3 Rinse column
x Place beaker beneath the column outlet.
x Switch on high-pressure pump and rinse column with eluent
for ca. 10 min, then switch off pump.
4 Connect column to detector block
x Screw outlet end of separating column 66
66
66
66 to inlet capillary 37
37
37
37
or suppressor inlet capillary 75
75
75
75.
Changing the separation system
If you wish to replace an IC separating column by a column of a differ-
ent type, proceed as follows (see Fig. 12 and Fig. 14):
1 Remove old column
x Switch off high-pressure pump and wait for pressure to drop.
x Unscrew column 66
66
66
66 from inlet capillary 37
37
37
37 of the detector
block or from the suppressor inlet capillary 75
75
75
75.
x Unscrew column 66
66
66
66 from column connection capillary 20
20
20
20 or
the precolumn.
2 Rinse with eluent
x Place beaker beneath the column connection capillary 20
20
20
20.
x Rinse IC system with eluent used for the separating column
(flow rate 1 mL/min) for ca. 15 min.
3 Connect new column to injector
x Remove end caps from column 66
66
66
66.
x Screw inlet end of separating column 66
66
66
66 (note flow direction)
to column connection capillary 20
20
20
20 or to the precolumn (see
section 2.7.7/section 2.7.8).
4 Rinse column
x Place beaker beneath the column outlet.
x Set flow rate for new separating column.
x Switch on high-pressure pump and rinse column with eluent
for ca. 10 min, then switch off pump.
5 Connect column to detector block
x Screw outlet end of separating column 66
66
66
66 to inlet capillary 37
37
37
37
or suppressor inlet capillary 75
75
75
75.