2.8 Suppressor module

792 Basic IC

39

4 Connect aspirating tubing for H

2

O

x Prepare aspirating tubing 68

68

68

68: Cut a piece of the 6.1803.020

PTFE tubing to the required length (normally ca. 120 cm).

x Insert one end of the aspirating tubing 68

68

68

68 into the rinsing

solution storage bottle and tighten it so that the tubing is

firmly held.

x Insert the free end of aspirating tubing 68

68

68

68 into one of the

openings 5

555 of the 792 Basic IC (see Fig. 2) and pull it suffi-

ciently far into the interior.

x Mount a coupling 70

70

70

70 (6.2744.030) to the inlet end of the front

pump tubing 72

72

72

72.

x Mount a 6.2744.010 Compression fitting at the end of the

aspirating tubing 68

68

68

68 and screw this compression fitting on to

the coupling 70

70

70

70 (see Fig. 14).

2.8.4

Connection of the suppressor module

The three inlets and outlets numbered 1...3 on the suppressor module

39

39

39

39 each have 2 permanently mounted PTFE capillaries, which must be

connected as follows (see Fig. 14 and Fig. 15):

1 Inlet capillary for eluent

x Screw inlet capillary 75

75

75

75 marked with "Eluent" of suppressor

module 39

39

39

39 to outlet end of separating column 66

66

66

66 using a

6.2744.010 Compression fitting.

2 Outlet capillary for eluent

x Screw outlet capillary 76

76

76

76 marked with "Detector" of suppres-

sor module 39

39

39

39 to coupling 26

26

26

26 using a 6.2744.010 Compres-

sion fitting.

x Screw inlet capillary 37

37

37

37 of detector block 38

38

38

38 to other end of

coupling 26

26

26

26.

3 Inlet capillary for H

2

SO

4

x Attach inlet capillary 78

78

78

78 marked with "H

2

SO

4

" of suppressor

module 39

39

39

39 using a 6.2744.010 Compression fitting to the filter

unit PEEK 28

28

28

28 connected to the rear pump tubing 71

71

71

71.

4 Outlet capillary for H

2

SO

4

x Pull outlet capillary 80

80

80

80 marked with "Waste" of the suppressor

module 39

39

39

39 from below through one of the openings 6666 out of

the inner compartment of the 792 Basic IC.

x Lead outlet capillary 80

80

80

80 to a sufficiently large waste container

and fix it in place.